Dose Estimation of Patient and Operator in Swallowing CT Examination Performed with a 320-detector-Row Multislice CT

Education Exhibits

Presented in 2014

Participants

Masanao Kobayashi PhD, RT, Presenter: Nothing to Disclose
Yasuki Asada PhD, Abstract Co-Author: Nothing to Disclose
Kosuke Matsubara PhD, Abstract Co-Author: Nothing to Disclose
Yuta Matsunaga RT, Abstract Co-Author: Nothing to Disclose
Ai Kawaguchi RT, Abstract Co-Author: Nothing to Disclose
Kazuhiro Katada MD, Abstract Co-Author: Consultant, Toshiba Corporation
Hiroshi Toyama, Abstract Co-Author: Nothing to Disclose
Kichiro Koshida PhD, Abstract Co-Author: Nothing to Disclose
Shouichi Suzuki PhD, Abstract Co-Author: Nothing to Disclose

TEACHING POINTS

Recently, attempts to develop new types of swallowing function analysis with 320-detector-row multislice CT (320-MDCT) have been reported. The present report addresses (1) patient exposure, (2) operator exposure, and (3) spatial dose distribution. For dose measurement, a human-body phantom in which 303 thermoluminescent dosimeter elements were inserted and a survey meter was used. Swallowing CT (SCT) was performed at 120kV, 10mA, 0.35sec/rot, 160mm/rot, and tilt angle of 22 (volume CT dose index displayed on the console 0.8 mGy, dose length product 12.1 mGy cm). The effective dose for the patient was 3.9 mSv. The conversion factor for obtaining the effective dose was 0.0066 mSv/mGy cm. The effective dose for the operator was 0.002 mSv. In the operator exposure measurement, the ambient dose equivalent H*(10), that would be produced by an expanded and aligned radiation field at a depth 10 mm in the International Commission on Radiation Units and Measurements sphere, was 0.012 mSv. In this report, the safety of SCT, which has become possible with the introduction of 320-MDCT, was evaluated by measurement of the exposure to the patient and operator.

TABLE OF CONTENTS/OUTLINE

Radiation protection

PDF UPLOAD

http://abstract.rsna.org/uploads/2014/14014312/14014312_58vr.pdf

Cite This Abstract

Kobayashi, M, Asada, Y, Matsubara, K, Matsunaga, Y, Kawaguchi, A, Katada, K, Toyama, H, Koshida, K, Suzuki, S, Dose Estimation of Patient and Operator in Swallowing CT Examination Performed with a 320-detector-Row Multislice CT.  Radiological Society of North America 2014 Scientific Assembly and Annual Meeting, - ,Chicago IL. http://archive.rsna.org/2014/14014312.html