Abstract Archives of the RSNA, 2013
SSE19-02
QIBA2 FDG PET Reading Study: Variability of Liver FDG Uptake Measurements across Different Sites
Scientific Formal (Paper) Presentations
Presented on December 2, 2013
Presented as part of SSE19: Nuclear Medicine (Quantitative Imaging)
Joo Hyun O MD, Abstract Co-Author: Nothing to Disclose
Edward A. Eikman MD, Abstract Co-Author: Nothing to Disclose
Jaime L Montilla-Soler,, Abstract Co-Author: Nothing to Disclose
Paul Eugene Kinahan PhD, Abstract Co-Author: Research Grant, General Electric Company
Co-founder, PET/X LLC
James Michael Mountz MD, PhD, Abstract Co-Author: Nothing to Disclose
Eric Perlman, Abstract Co-Author: Nothing to Disclose
John Sunderland PhD, Abstract Co-Author: Nothing to Disclose
Heather Jacene MD, Abstract Co-Author: Nothing to Disclose
Nathan C. Hall MD, PhD, Abstract Co-Author: Consultant, Enlyton, Ltd
Michael Vinzenz Knopp MD, PhD, Abstract Co-Author: Nothing to Disclose
Abdel K. Tahari MD, PhD, Abstract Co-Author: Nothing to Disclose
Ronald Boellaard PhD, Abstract Co-Author: Nothing to Disclose
Otto S. Hoekstra MD, Abstract Co-Author: Nothing to Disclose
Li Huo, Abstract Co-Author: Nothing to Disclose
Hye Ok Kim, Abstract Co-Author: Nothing to Disclose
Sun Young Chae, Abstract Co-Author: Nothing to Disclose
Sae Jung Na, Abstract Co-Author: Nothing to Disclose
Sung Hoon Kim, Abstract Co-Author: Nothing to Disclose
Mike Sathekge, Abstract Co-Author: Nothing to Disclose
Moses Modiselle, Abstract Co-Author: Nothing to Disclose
Sally Barrington, Abstract Co-Author: Nothing to Disclose
Andrew Mark Scott, Abstract Co-Author: Nothing to Disclose
Sam Berlangieri, Abstract Co-Author: Nothing to Disclose
Andrew Quon MD, Abstract Co-Author: Speakers Bureau, Eli Lilly and Company
Research Consultant, Koninklijke Philips Electronics NV
Jeffrey P. Leal BA, Abstract Co-Author: Nothing to Disclose
Muhammad Ali Chaudhry MD,MBBS, Abstract Co-Author: Nothing to Disclose
Richard L. Wahl MD, Presenter: Patents, Naviscan, Inc
Patents, GlaxoSmithKline plc
Patents, Spectrum Pharmaceuticals, Inc
Research Consultant, GlaxoSmithKline plc
Research Consultant, Nihon Medi-Physics Co, Ltd
Research support, General Electric Company
Research support, Molecular Insight Pharmaceuticals, Inc
Research support, Cell Point LLC
To determine the variability in measurement of FDG uptake in normal liver background in identical human PET studies analyzed by differing readers, software and performance sites.
Baseline and post-therapy FDG PET/CT images of a single patient were distributed digitally to 15 sites in North America, Europe, Africa, Asia and Australia in an IRB approved study. Readers at each site measured the background activity by placing a 3 cm diameter spherical volume of interest (VOI) in right side of the liver. Workstation software from 8 different vendors were used for the quantification. The standard uptake value (SUV) and SUV corrected for lean body mass (SUL) were measured using local standard approaches.
The liver background SUV ranged from 1.47 to 2.19 at baseline with standard deviation (SD) of 0.16; and ranged from 1.80 to 2.96 (SD 0.26) at follow-up. The liver SUL measurements ranged from 0.91 to 1.56 (SD 0.12) at baseline; and from 1.20 to 2.02 (SD 0.17) at follow-up.
SUV and SUL measurements of the normal liver derived from the same set of FDG PET/CT images showed generally similar, but not consistently the same, values. The cause of the variance of liver measurements is likely due to both human VOI selection methods and software differences.
These results indicate that varying quantitative results can be extracted from identical PET/CT images and suggest the need to more fully standardize the PET analytic process.
O, J,
Eikman, E,
Montilla-Soler,, J,
Kinahan, P,
Mountz, J,
Perlman, E,
Sunderland, J,
Jacene, H,
Hall, N,
Knopp, M,
Tahari, A,
Boellaard, R,
Hoekstra, O,
Huo, L,
Kim, H,
Chae, S,
Na, S,
Kim, S,
Sathekge, M,
Modiselle, M,
Barrington, S,
Scott, A,
Berlangieri, S,
Quon, A,
Leal, J,
Chaudhry, M,
Wahl, R,
QIBA2 FDG PET Reading Study: Variability of Liver FDG Uptake Measurements across Different Sites. Radiological Society of North America 2013 Scientific Assembly and Annual Meeting, December 1 - December 6, 2013 ,Chicago IL.
http://archive.rsna.org/2013/13024995.html