Simulated and Measured Analyses of Spheroprobability and Three Dimensional Setup Error
Scientific Papers
Presented on November 26, 2006
Presented as part of SSA18: Physics (Radiation Therapy)
Todd Jefferson Scarbrough MD, Presenter: Nothing to Disclose
Joseph Y Ting PhD, Abstract Co-Author: Nothing to Disclose
Samuel J Wang MD,PHD, Abstract Co-Author: Nothing to Disclose
Clifton David Fuller BS, Abstract Co-Author: Nothing to Disclose
Charles Richard Thomas MD, Abstract Co-Author: Nothing to Disclose
Unidimensionally, setup error clusters near the isocenter and follows a normal distribution. This has been expressed with the aphorism: "Typical patient setup error is about plus or minus a half centimeter." We examined the three-dimensional (3D) implications of "typical setup error" via large simulated and measured datasets.
We simulated 10000 patient setups by generating 30000 normally distributed lateral, vertical, and longitudinal setup points with each dimension having a mean of ~0 mm and a std dev of ~5 mm. We also measured (via kV X-ray analysis of gold marker fiducials within the prostate after setup to skin marks) 1019 separate patient setups. We compared the simulated and measured data.
Patients are almost never set up very near the isocenter in 3D space (0.02% of simulated setups were within 1 mm of the isocenter in 3D space). The simulated data reveal that if average unidimensional setup error has a mean of ~0 mm and std dev of 5 mm, average 3D setup error is 8.00 mm +/- 4.0 mm. Our measured data reveal average setup error to be 7.98 mm +/- 3.5 mm, not significantly disagreeing with the simulated data (P=0.86). Only 24% of all measured patient setups were within 5 mm of the isocenter in 3D space; 76% of the time, patients are set up more than 5 mm from the planned isocenter. Setups can be plotted in 3D space via spheroprobability plots..
Average setup error is ~8 mm; this error is probably underappreciated by infrequent orthogonal films. Patients are uncommonly set up within 5 mm of the planned isocenter (patients thus are commonly in need of setup error correction). Intermittent port film analysis will detect shifts, but the magnitude of setup error in 3D space will always be greater than the error witnessed in 1D or 2D views. 3D setup error can be represented using spheroprobability plots, and spheroprobability plots can be used to determine PTV margins.