Three Dimensional Setup Errors for Prostate Radiotherapy Verified Using Cone-beam CT
Scientific Papers
Presented on November 26, 2006
Presented as part of SSA18: Physics (Radiation Therapy)
Sua Yoo, Presenter: Nothing to Disclose
Accurate patient setup is essential for radiation treatment in accordance with the treatment plan. In this work, we investigate the extent of mismatch of 3D bony anatomy and the extent of mismatch of 3D soft tissue after making set-up error corrections based on 2D bony anatomy in orthogonal radiographs.
A linear accelerator (Varian 21EX, Varian Medical Systems) equipped with an on-board imaging (OBI) system enables 2D kilovoltage (kV) radiographic imaging and 3D cone beam computed tomography (CBCT). CBCT is reconstructed using OBI projections acquired over an entire gantry rotation. Prior to patient treatment, kV orthogonal images were taken and compared to the planning digitally-reconstructed radiographs (DRR) using 2D matching tool. CBCT is used to provide additional definition of 3D bony anatomy and soft tissues. The CBCT image was superimposed on the planning CT image, and the couch shift required to correct 3D set-up errors was assessed using 3D matching. For 16 prostate cancer patients, kV orthogonal images were taken and 2D matching was performed to determine the set-up error based on matching 2D bony anatomy. The couch shift to correct set-up error determined from 2D matching was applied. Following the couch shift, CBCT was acquired. 3D matching was performed to determine the residual set-up error to match 3D bony anatomy and to separately determine the set-up error to match the soft tissue (e.g., the prostate). The resulting couch shifts were not actually applied but were recorded for comparison. This procedure was followed daily for the 16 patients, daily on the first week and weekly thereafter.
The figure shows the average required shifts obtained from matching of 2D bony anatomy, 3D bony anatomy, and 3D soft tissue in all directions and magnitude. The residual set-up error is negligible for 3D bony anatomy after corrections of set-up error for 2D bony anatomy. However, there were large set-up errors (average 0.6cm in magnitude) based on 3D soft tissue matching, even after matching of 2D bony anatomy. These errors are most likely due to the large variations in the bladder and rectum since the time the planning CT was acquired.
To achieve a good treatment outcome it is imperative that the target be accurately localized as originally planned. Our study using CBCT shows that fairly sizeable errors remain in the prostate target location after 2D orthogonal radiograph-based shifts. Thus, prostate CBCT imaging provides valuable soft-tissue information that is complementary to conventionally employed orthogonal radiographs.
Yoo, S,
Three Dimensional Setup Errors for Prostate Radiotherapy Verified Using Cone-beam CT. Radiological Society of North America 2006 Scientific Assembly and Annual Meeting, November 26 - December 1, 2006 ,Chicago IL.