Abstract Archives of the RSNA, 2006
SSA18-01
Automated Method for Measurement of Displacement Vectors of Lung Tumor Position on EPID Portal Cine Images without Implanted Markers in Stereotactic Radiotherapy
Scientific Papers
Presented on November 26, 2006
Presented as part of SSA18: Physics (Radiation Therapy)
Hidetaka Arimura PhD, Presenter: Nothing to Disclose
Shigeo Anai BA, Abstract Co-Author: Nothing to Disclose
Satoshi Yoshidome BA, Abstract Co-Author: Nothing to Disclose
Katsumasa Nakamura MD, Abstract Co-Author: Nothing to Disclose
Yoshiyuki Shioyama, Abstract Co-Author: Nothing to Disclose
Hiromi Terashima MD,PHD, Abstract Co-Author: Nothing to Disclose
Yoshihiko Onizuka, Abstract Co-Author: Nothing to Disclose
et al, Abstract Co-Author: Nothing to Disclose
To develop an automated method for measurement of the displacement vectors of lung tumor position on portal cine images acquired by an electric portal imaging device (EPID) without implanted markers in stereotactic radiotherapy.
The displacement vector of a tumor position was derived by using a template matching between the first cine image (reference image) and each consecutive cine image in a treatment for a patient. First, an irradiation field determined by the MLC was extracted in each cine image by using an automated thresholding technique based on gray-level histogram analysis. Second, each cine image was cropped by a rectangular field including the irradiation field. Third, the surrounding area of the MLC field was filled with an average pixel value in the irradiation field (referred to as MLC image), and a Gaussian filter was applied to the MLC image for smoothing the edge of the MLC and reducing background noise. Finally, the displacement vector of the tumor on each cine image was determined, based on the coordinate system in the reference image, from the position in which took the maximum correlation coefficient between the first cine image and the consecutive image. Cine images of the EPID with 512×384 pixels (pixel size:0.78mm) and a sampling rate of 0.5 frame/sec were acquired from 5 patients with non-small cell lung cancer. For validation of our method, we calculated the inner products (correlation value) between displacement vectors calculated by our method and those obtained manually by two radiation oncologists.
Tumor displacements in all cine images for three cases calculated by our method correlated well with those measured by the oncologists (r=0.847, 0.819, 0.715), but correlation values for two cases were lower.
This preliminary result suggested that our computerized method based on the EPID cine images would be useful for measurement of the displacement vectors of lung tumor position without implanted markers.
Our method without implanted markers may be equal to other methods with implanted markers and could be recommended in quality assurance of the stereotactic radiotherapy.
Arimura, H,
Anai, S,
Yoshidome, S,
Nakamura, K,
Shioyama, Y,
Terashima, H,
Onizuka, Y,
et al, ,
Automated Method for Measurement of Displacement Vectors of Lung Tumor Position on EPID Portal Cine Images without Implanted Markers in Stereotactic Radiotherapy. Radiological Society of North America 2006 Scientific Assembly and Annual Meeting, November 26 - December 1, 2006 ,Chicago IL.
http://archive.rsna.org/2006/4428428.html