To investigate the impact of iodine concentration levels on absorbed radiation dose in blood for various incident photon energies in a cardiac CT model.
METHOD AND MATERIALSA representative model for cardiac CT imaging was defined, including a cardiac-chest model and the exposure geometry of a wide-beam CT scanner (Revolution CT, GE Healthcare). In the chest model, the heart was represented by a volume filled with blood including six relevant iodine contrast fractions (c = 0-50, step 10 mg I/mL), surrounded by air (lungs) and water (soft tissue). A one-heartbeat CT acquisition was simulated by 8 rotational projections over 360° with 7 types of photon spectra: X-ray CT spectra 70, 80, 100, 120, 140 kVp and gamma spectra 137Cs (662 keV) and 60Co (1250 keV). For each energy-contrast combination, the absorbed dose (D), was calculated using the f6 tally of MCNPX code, with enough particles to maintain the uncertainties below 1%.
RESULTSWith X-ray energies, the presence of iodine in blood increased D between 70 and 140 kVp by 1.7 (70 kVp, c = 10 mg I/ml) and 2.9 (140 kVp, c = 50 mg I/ml) fold, following an exponential trend.On the other hand, for 662 and 1250 keV gamma energies, a negligible increase of less than 4% and 1% was observed, respectively, even for c = 50 mg I/ml. This underlines the fact that the elevated D with X-ray energies is caused by the photoelectric effect.
CONCLUSIONOur results indicate that the presence of iodine increases radiation dose in CT due to increased photoelectric effect, and that there is an exponential trend with iodine concentration.
CLINICAL RELEVANCE/APPLICATIONA reduction in iodine contrast administration load also reduces radiation dose to the patient.